Phase-contrast imaging is a method of imaging that has a range of different applications. It exploits differences in the refractive index of different materials to differentiate between structures under analysis. In conventional light microscopy, phase contrast can be employed to distinguish between structures of similar transparency, and to examine crystals on the basis of their double refraction. This has uses in biological, medical and geological science. In X-ray tomography, the same physical principles can be used to increase image contrast by highlighting small details of differing refractive index within structures that are otherwise uniform. In transmission electron microscopy (TEM), phase contrast enables very high resolution (HR) imaging, making it possible to distinguish features a few Angstrom apart (at this point highest resolution is 60 pm).
See also: Phase contrast microscopy and Quantitative phase contrast microscopy
Phase contrast takes advantage of the fact that different structures have different refractive indices, and so bend light and delay its passage through the sample by different amounts. The retardation of the light results in some waves being 'out of phase' with others, and so to the human eye a microscope in phase contrast mode effectively darkens or brightens particular areas to reflect this change.
Phase contrast is used extensively in optical microscopy, in both biological and geological sciences. In biology, it is employed in viewing unstained biological samples with the human eye, making it possible to distinguish between structures that are of very similar transparency.
In geology, phase contrast is exploited in a different way to highlight differences between mineral crystals cut to a standardised thin section (usually 30 µm) and mounted under a light microscope. Crystalline materials are capable of exhibiting double refraction, in which light rays entering a crystal are split into two beams that may exhibit different refractive indices, depending on the angle at which they enter the crystal. The phase contrast between the two rays can be detected with the human eye using particular optical filters. As the exact nature of the double refraction varies for different crystal structures, phase contrast aids in the identification of minerals.
There are four main techniques for x-ray phase-contrast imaging, which use different principles to convert phase variations in the x-rays emerging from the object into intensity variations at an x-ray detector.[1][2] Propagation-based phase contrast[3] uses free-space propagation to get edge enhancement, talbot and polychromatic far-field interferometry[2][4] uses a set of diffraction gratings to measure the derivative of the phase, refraction-enhanced imaging[5] uses an analyzer crystal also for differential measurement, and x-ray interferometry[6] uses a crystal interferometer to measure the phase directly. The advantage of these methods compared to normal absorption-contrast x-ray imaging is higher contrast that makes it possible to see smaller details. One disadvantage is that these methods require more sophisticated equipment, such as synchrotron or microfocus x-ray sources, x-ray optics, and high resolution x-ray detectors. This sophisticated equipment provides the sensitivity required to differentiate between small variations in the refractive index of x-rays passing through different media. The refractive index is normally smaller than 1 with a difference from 1 between 6993100000000000000♠10−7 and 6994100000000000000♠10−6.
All of these methods produce images that can be used to calculate the projections (integrals) of the refractive index in the imaging direction. For propagation-based phase contrast there are phase-retrieval algorithms, for talbot interferometry and refraction-enhanced imaging the image is integrated in the proper direction, and for x-ray interferometry phase unwrapping is performed. For this reason they are well suited for tomography, i.e. reconstruction of a 3D-map of the refractive index of the object from many images at slightly different angles. For x-ray radiation the difference from 1 of the refractive index is essentially proportional to the density of the material.
Synchrotron X-ray tomography can employ phase contrast imaging to enable imaging of the interior surfaces of objects. In this context, phase contrast imaging is used to enhance the contrast that would normally be possible from conventional radiographic imaging. A difference in the refractive index between a detail and its surroundings causes a phase shift between the light wave that travels through the detail and that which travels outside the detail. An interference pattern results, marking out the detail.[7]
This method has been used to image Precambrian metazoan embryos from the Doushantuo Formation in China, allowing the internal structure of delicate microfossils to be imaged without destroying the original specimen.[8]
This section has an unclear citation style. The references used may be made clearer with a different or consistent style of citation, footnoting, or external linking. The citations lack any details besides author and title. (July 2013) (Learn how and when to remove this template message) |
In the field of transmission electron microscopy, phase-contrast imaging may be employed to image columns of individual atoms. This ability arises from the fact that the atoms in a material diffract electrons as the electrons pass through them (the relative phases of the electrons change upon transmission through the sample), causing diffraction contrast in addition to the already present contrast in the transmitted beam. Phase-contrast imaging is the highest resolution imaging technique ever developed, and can allow for resolutions of less than one angstrom (less than 0.1 nanometres). It thus enables the direct viewing of columns of atoms in a crystalline material.[9][10]
The interpretation of phase-contrast images is not a straightforward task. Deconvolving the contrast seen in an HR image to determine which features are due to which atoms in the material can rarely, if ever, be done by eye. Instead, because the combination of contrasts due to multiple diffracting elements and planes and the transmitted beam is complex, computer simulations are used to determine what sort of contrast different structures may produce in a phase-contrast image. Thus, a reasonable amount of information about the sample needs to be understood before a phase contrast image can be properly interpreted, such as a conjecture as to what crystal structure the material has.
Phase-contrast images are formed by removing the objective aperture entirely or by using a very large objective aperture. This ensures that not only the transmitted beam, but also the diffracted ones are allowed to contribute to the image. Instruments that are specifically designed for phase-contrast imaging are often called HRTEMs (high resolution transmission electron microscopes), and differ from analytical TEMs mainly in the design of the electron beam column. Whereas analytical TEMs employ additional detectors attached to the column for spectroscopic measurements, HRTEMs have little or no additional attachments so as to ensure a uniform electromagnetic environment all the way down the column for each beam leaving the sample (transmitted and diffracted). Because phase-contrast imaging relies on differences in phase between electrons leaving the sample, any additional phase shifts that occur between the sample and the viewing screen can make the image impossible to interpret. Thus, a very low degree of lens aberration is also a requirement for HRTEMs, and advances in spherical aberration (Cs) correction have enabled a new generation of HRTEMs to reach resolutions once thought impossible.