Countercurrent chromatography (CCC) is a liquid chromatography technique that uses two immiscible liquid phases and no solid support.[1][2] One liquid acts as the stationary phase and the other as the mobile phase. The liquid stationary phase is held in place by gravity or by centrifugal force.
Standard column chromatography uses a solid stationary phase and a liquid mobile phase, while gas chromatography (GC) uses a solid or liquid stationary phase on a solid support and a gaseous mobile phase. By contrast, in liquid-liquid chromatography, both the mobile and stationary phases are liquid. By eliminating solid supports, permanent adsorption of the analyte onto the column is avoided, and a high recovery of the analyte can be achieved.[3] The instrument is also easily switched between normal-phase and reversed-phase modes of operation simply by changing the mobile and stationary phases. With liquid chromatography, operation is limited by the composition of the columns and media commercially available for the instrument. Nearly any pair of immiscible solutions can be used in liquid-liquid chromatography provided that the stationary phase can be successfully retained.
Solvent costs are also generally lower than for high-performance liquid chromatography (HPLC), and the cost of purchasing and disposing of solid adsorbents is eliminated. Another advantage is that experiments conducted in the laboratory can be scaled to industrial volumes. When GC or HPLC is carried out with large volumes, resolution is lost due to issues with surface-to-volume ratios and flow dynamics; this is avoided when both phases are liquid.[4]
The CCC separation process can be thought of as occurring in three stages: mixing, settling, and separation of the two phases (although they often occur continuously). Vigorous mixing of the phases is critical so that the interface between them has a large area, and the analyte can distribute between the phases according to its partition coefficient.
The mobile phase mixes with then settles from the stationary phase throughout the column. The degree of stationary phase retention (inversely proportional to the amount of stationary phase loss or "bleed" in the course of a separation) is a crucial parameter. Common factors that influence stationary phase retention are flow rate, solvent composition of the biphasic solvent system, and the G-force created by rotation. The settling time is a property of the solvent system and the sample matrix, both of which greatly influence stationary phase retention.[5]
Droplet CCC is the oldest form of CCC.[1] It uses only gravity to move the mobile phase through the stationary phase. In descending mode, droplets of the denser mobile phase and sample are allowed to fall through a column of the lighter stationary phase using only gravity.
If a less dense mobile phase is used it will rise through the stationary phase; this is called ascending mode. The eluent from one column is transferred to another; the more columns that are used, the more theoretical plates can be achieved. The disadvantage of DCCC is that flow rates are low, and poor mixing is achieved for most binary solvent systems, which makes this technique both time-consuming and inflexible.
The modern era of CCC began with the development by Dr. Yoichiro Ito of the planetary centrifuge and the many possible column geometries it can support.[6] These devices make use of a little-known means of making non-rotating connections between the stator and the rotor of a centrifuge. (It is beyond the scope of this discussion to describe the method of accomplishing this. Any of the several books available on CCC discuss it thoroughly.)[1][7] [8] [9] [10] [11] [12]
Functionally, the high-speed CCC apparatus consists of a helical coil of inert tubing which rotates on its planetary axis and simultaneously rotates eccentrically about another solar axis. (These axes can be made to coincide, but the most common or type J CCC is discussed here.) The effect is to create zones of mixing and zones of settling which progress along the helical coil rapidly. This produces a highly favorable environment for chromatography.
There are numerous potential variants upon this instrument design. The most significant of these is the toroidal CCC. This instrument does not employ planetary motion. In some respects it is very like CPC, but retains the advantage of not needing rotary seals. It also employs a capillary tube instead of the larger-diameter tubes employed in the helices of the other CCC models. This capillary passage makes the mixing of two phases very thorough, despite the lack of shaking or other mixing forces. This instrument provides rapid analytical-scale separations, which can nonetheless be scaled up to either of the larger-scale CCC instruments. See, for instance, the Xanthanolide purification found in.[13]
The operating principle of CCC equipment requires a column consisting of a tube coiled around a bobbin. The bobbin is rotated in a double-axis gyratory motion (a cardioid), which causes a variable gravity (G) field to act on the column during each rotation. This motion causes the column to see one partitioning step per revolution and components of the sample separate in the column due to their partitioning coefficient between the two immiscible liquid phases used.
HPCCC works in much the same way as HSCCC but with one vital difference. A seven-year R&D process that has produced HPCCC instruments that generated 240 G's, compared to the 80 G's of the HSCCC machines. This increase in G-level and larger bore of the column has enabled a tenfold increase in through put, due to improved mobile phase flow rates and a much higher stationary phase retention.[14]
Countercurrent chromatography is a preparative liquid chromatography technique, however with the advent of the higher-G HPCCC instruments it is now possible to operate instruments with sample loadings as low as a few milligrams, whereas in the past 100s of milligrams had been necessary.
Major application areas for this technique include natural products purification and also drug development.
Centrifugal partition chromatography (CPC) was invented in the eighties by the Japanese company Sanki Engineering Ltd, whose president was the late Kanichi Nunogaki.[15] CPC has been extensively developed in France starting from the late nineties. CPC uses centrifugal force to speed separation and achieves higher flow rates than DCCC (which relies on gravity). This technique is sometimes sold under the name FCPC or SCPC.
This technique is still based on the principles of liquid/liquid partitioning chromatography: two non-miscible liquid phases are mixed together to form a two-phase system, and are then separated multiple times.[16] The individual solutes are isolated based on the different partitioning coefficients of each compound in this two-phase system. One of the liquid phases of the two-phase system is used as a stationary liquid phase: it is fed into the column (the rotor) while the latter is spinning at moderate rotational speed. The stationary phase is retained inside the rotor by the centrifugal force generated. The second phase of the two-phase system is used as the mobile phase containing the solutes to be extracted. It is fed under pressure into the rotor and pumped through the stationary phase. Both phases are mixed together. It is at that time that the exchange of molecules between the two phases occurs. The separation of the solutes is achieved as a function of the specific partitioning coefficient (Kd) of each solute between the mobile and stationary phases. The mobile phase then decants at each cell outlet thus entering the next cell. The eluted fractions of the mobile and stationary phases are collected over a period of several minutes to several hours. These fractions, or eluates, will contain the individual purified solutes.
The centrifugal partition chromatograph is constituted with a unique rotor (=column). This rotor rotates on its central axis (while HSCCC column rotates on its planetary axis and simultaneously rotates eccentrically about another solar axis). With less vibrations and noise, the CPC offers a wider rotation speed range (from 500 to 2000 rpm) than HSCCC. That allows a better decantation and retention for unstable biphasic system (e.g., aqueous aqueous systems or butanol/water systems).
The CPC rotor is constituted by the superposition of disks engraved with small cells connected by head / tail ducts. These cells, where the chromatographic separation takes place, can be compared to lined-up separate funnels.
Rotor (column)
Twin Cell Disc
Detail of Twin Cell disc
Detail of Twin Cell
The rotor is filled with the stationary phase, which stays inside the rotor thanks to the rotation speed, while the mobile phase is pumped through. CPC can be operated in either descending or ascending mode, where the direction is relative to the force generated by the rotor rather than gravity. According to the fast and permanent evolution of the cells design, the efficiency and flow rate with low back pressure are improved.
Ascending mode
Descending mode
The CPC offers now the direct scale up from the analytical apparatuses (few milliliters) to industrial apparatuses (some liters) for fast batch production.
|